Increased modality choice with a citrate based anticoagulant, coupled with the ability to swap between heparin and citrate utilising the same line set are key highlights of . Whilst maintaining the automatic titration, unique fluid balance auto-compensation, familiar interface and easy trouble-shooting as appreciated on the original AquariusTM.
Citrate solution is infused into the blood circuit before the blood pump. To compensate for the amount of Calcium lost in the filtrate, Calcium solution is infused into the blood circuit just before the air detector system and after the drip chamber.
Citrate containing replacement solution is infused into the blood circuit after the blood pump and before the hemofilter. Calcium solution is infused into the blood circuit just before the air detector system and after the drip chamber.
The AquariusTM system with RCA system provides complete prescription flexibility with one Aqualine RCA for all therapies, regardless of anticoagulant choice. Regular Aqualine can also be used with the AquariusTM system with RCA for treatments without citrate anticoagulation.
Products | CITRASET RCA 12 | CITRASET RCA 19 |
---|---|---|
Aquamax ® filter | HF12 | HF19 |
Bloodline | Aqualine RCA | Aqualine RCA |
Bloodline blood volume | 100 ml | 100 ml |
Ancillaries | 1 effluent bag 5 liter + 1 empty priming bag 2 liter + 1 two-way connector |
Flow rates | ADULT | |
---|---|---|
Blood pump | TPE CVVH |
30 to 250 ml / min 30 to 300 ml / min |
Post-dilution pump | TPE CVVH |
0 or 500 to 3,000 ml / h 0 or 500 to 6,000 ml / h |
Citrate pump | 0 to 650 ml / h | |
Calcium pump | 0 or 2 to 300 ml / h | |
Fluid loss | CVVH | 0 to 2,000 ml / h |
Scales | ||
Citrate / Calcium maximum load | 2.2 kg | |
Citrate and Calcium intended use in CVVH Post-dilution | ||
Concentration of citrate | between 110 mmol / l and 150 mmol / l | |
Concentration of calcium supplementation | approx. 5 mmol / l to 20 mmol / l |